Features | Z540.1 Compliant with OOT Data | Z540.1 Comp with Data | 17025:2017 Accredited (Simple Decision) |
17025:2017 Accredited (with Guard Banding) |
---|---|---|---|---|
CalCloud™ Database Access | ✔ | ✔ | ✔ | ✔ |
Calibrated to a Defined Specification. | ✔ | ✔ | ✔ | ✔ |
Ensure Test Accuracy Requirements | ✔ | ✔ | ✔ | ✔ |
Equipment Traceability to an (SI) maintained by NMI’s | ✔ | ✔ | ✔ | ✔ |
NAVSEA 009-04 * | ✔ | ✔ | ✔ | ✔ |
MIL-STD-45662A | ✔ | ✔ | ✔ | ✔ |
Supplemental Data | ✔ | ✔ | ✔ | |
Covered by Laboratory Scope of Accreditation | ✔ | ✔ | ||
Stated Risk Contributions, Metrological Traceability, and Measurement Uncertainty | ✔ | ✔ | ||
Reduced Risk Through Guardbanding ** | ✔ | |||
Z540.3 Accredited Service | ✔ | |||
NAVSEA 009-04 *** | ✔ |
Z540.1 Compliant with OOT Data
Compliant calibration to ANSI/NCSL Z540.1-1994 provides complete coverage for recommended data range, out-of-tolerance results provided where applicable.
Z540.1 Compliant with Data
Compliant calibration to ANSI/NCSL Z540.1-1994 and includes the entire suite of data observed during the calibration.
17025:2017 Accredited (Simple Decision)
An ISO/IEC 17025:2017 accredited service which includes the entire suite of data observed during the calibration, the stated test uncertainty contributions identified in the calibration, and a defined ratio for the impact of uncertainties on measurements taken.
17025:2017 Accredited (with Guard Banding)
AN ISO/IEC 17025:2017 and/or ANSI/NCSL Z540.3-2006 accredited service which includes the entire suite of data observed during the calibration, the stated test uncertainty contributions identified in the calibration, a defined ratio for the impact of uncertainties on measurements taken, and tolerance limits which include the identified uncertainty impact identified to ensure a probability of false acceptance rate less than two percent.
*Standard Calibrations Inc. provides service which comply with NAVSEA Standard Item 009-04 to FY22 requirements. All measurement disciplines covered by Scope of Accreditation.
**Standard Calibrations Inc. provides 17025:2017 accredited services with guard banding utilizing Z540.3 Method 6 to ensure a probability of false acceptance rate less than two percent.
***Standard Calibrations Inc. provides service which complies with the new NAVSEA Standard Item 009-04 FY23 requirements.